Skip to content
 |  Tyco Electronics Segment Indicator  |   | 
Tyco Electronics - Our commitment. Your advantage.
Visit www.tycoelectronics.com Corcom Products
Home > Series > PEM > L

L Series

Power Entry Module with RFI Power Line Filter for General or Medical Applications

UL Recognized
CSA Certified
VDE Approved

Catalog Data Sheet

L Series

The L series power entry modules are compact units that combine a multi-function power entry module and high performance RFI filtering capabilities. They are available with either a four-voltage selector or a DPST on/off switch. Both variations can be specified with North American or European fusing capabilities and are available in either flange or snap-in mounting.

These filters are UL recognized, CSA certified, and VDE approved. The L series modules offer a choice of filters for general or medical applications.

EDL Models

The RFI filter is for general purpose applications where line-to-line and line-to-ground noise must be controlled. The filter is designed to meet the very low leakage requirements of VDE portable equipment. They are available in three current ratings.

EHL Models

This medical filter provides susceptibility protection without the leakage current associated with line-to-ground capacitors. Designed to allow equipment to meet UL544 for patient care and non-patient care equipment, the EHL filter has a maximum leakage current of 2 μA at 120 VAC 60 Hz.

See Appendix C for more information on UL544 or UL2601 health care applications.

Note: When using the models with an IEC connector, remember that the leakage current of the companion line cord, GA400, is more significant than that of the filter - which may disqualify this line cord from use in patient care applications.

 

Electrical Schematic

EL Models (with Switch)

click on image for larger view

DL Models (with 4-voltage selection)

click on image for larger view

HL Models (with Switch

click on image for larger view

HL Models (with 4-voltage selection)

click on image for larger view

Note 1: Provisioni for dual European style fusing
Note 2: On/ Off switch present only with "S" suffix

Resistor location for reference only

Specifications

Maximum leakage current, each line to ground :
    EDL
Models
EHL
Models
  @ 120 VAC 60 Hz: 0.25 mA 2 µA
  @ 250 VAC 50 Hz
0.50 mA 5 µA
     

Hipot rating (one minute):

  line to ground 1500 VAC
  line to line 1450 VDC
     
Operating Frequency: 50 / 60 Hz
     
Operating voltages:
  1S & 1SC Suffix (fixed) 250 VAC Max.
  4 & 4C Suffix (selectable) 100, 120, 220, 240 VAC
     
Rated Current:
  2EDL & 2EHL 2A
  4EDL & 4EHL 4A
  6EL, 6EDL & 6EHL 6A
   
Fuse (not included):

Accepts one .25" x 1.25" fuse or two 5 x 20mm fuses

   
Switch:

Double-insulated, rated for 10,000 operations at full load. 51 Amp inrush capability

   
Terminals: .110 [2.79] except for switch
.187 [4.8] for switch
   
  † Filtered models only
Minimum insertion loss in dB:
Line to ground in 50 ohm circuit

Current Rating Frequency - MHz
.05 .15 1 5 10 30  

EDL Models
2A 6 14 24 40 45 50  
4A 2 8 18 32 38 45  
6A 1 6 17 31 37 45  
EHL Models
6A 3 8 15 18 18 18  

 
Line to line in 50 ohm circuit

Current Rating Frequency - MHz
.05 .5 1 3 5 10 30

EDL Models
2A 7 16 21 23 37 47 50
4A 6 14 18 23 26 45 47
6A 6 15 20 25 24 45 50
EHL Models
6A 4 14 20 28 32 38 42

 

Voltage Selection

To change selected voltage: disconnect the power cord; open cover using a small blade screwdriver or similar tool; insert the tool into the voltage selection slot and remove wheel from unit; select desired voltage; replace wheel into unit and close cover, making sure the selected voltage appears in connector window.

Click on image for larger version

Note: All models have North American or dual European fusing capability and are available in either flange or snap-in mounting.

 

Case Styles - Unfiltered

Unfiltered Flange Models

Typical Dimensions
Terminals (except for switch) .110 Switch Terminals .187
2.79 4.765
Mounting Holes (2)
.126 Dia  
3.20

Unfiltered Snap-In Models

Typical Dimensions
Terminals (except for switch) .110 Switch Terminals .187
2.79 4.765
Notes:

(1) For panel thickness of .031 - .079 [0.8 - 2.0]

(2) For panel thickness of .083 - .126 [2.1 - 3.2]

(3) For snap-in application, the 1.22 [28.5] sides of the cutout must have a .02 [.508] radius on the installation side

Snap-in models allow for front mounting only

 

Case Styles - Filtered

Filtered Flange Models

 

 

Typical Dimensions
Terminals (except for switch) .110 Switch Terminals .187
2.79 4.765
Mounting Holes (2)
.126 Dia  
3.20

 

Filtered Snap-In Models

 

3

Typical Dimensions
Terminals (except for switch) .110 Switch Terminals .187
2.79 4.765
Notes:

(1) For panel thickness of .031 - .079 [0.8 - 2.0]

(2) For panel thickness of .083 - .126 [2.1 - 3.2]

(3) For snap-in application, the 1.22 [28.5] sides of the cutout must have a .02 [.508] radius on the installation side

Snap-in models allow for front mounting only

Metric fuse models have additional jumper from filter to module

 

Case Dimensions

Part No. A B C D E F
(max.) ± .015 (max.) (max.) (max.) (max.)
± .381

Unfiltered
Flange Mount

1.98 1.575 2.3 2.14 1.66 1.11
50.29 40.0 58.42 54.36 42.16 28.19

Unfiltered
Snap-In

1.28   2.3 2.14 1.66 1.11
32.51   58.42 54.36 42.16 28.19

Filtered
Flange Mount

1.98 1.575 2.3 2.14 2.01 1.11
50.29 40.0 58.42 54.36 51.05 28.19
Unfiltered
Snap-In
1.28   2.3 2.14 2.01 1.11
32.51   58.42 54.36 51.05 28.19

Part Numbers

Click on any part number for distributor stock status and eCommerce Links. For more information refer to the Corcom Part Numbering System reference.

UNFILTERED Models

Part Number Current Rating Available Voltage Selection Positions DPST On/Off Switch Mounting Style Fuseholder Type

6EL1S 6A 1 Flange North American
6EL1SC 6A 1 Snap-In North American
6EL1SM 6A 1 Flange Metric
6EL1SCM 6A 1 Snap-In Metric
6EL4 6A 4   Flange North American
6EL4C 6A 4   Snap-In North American
6EL4M 6A 4   Flange Metric
6EL4CM 6A 4   Snap-In Metric



FILTERED Models

Part Number Filter Type Current Rating Available Voltage Selection Positions DPST On/Off Switch Mounting Style Fuseholder Type

2EDL1S Linear 2A 1 Flange North American
2EDL1SC Linear 2A 1 Snap-In North American
2EDL1SM Linear 2A 1 Flange Metric
2EDL1SCM Linear 2A 1 Snap-In Metric
2EDL4 Linear 2A 4 Flange North American
2EDL4C Linear 2A 4 Snap-In North American
2EDL4M Linear 2A 4 Flange Metric
2EDL4CM Linear 2A 4 Snap-In Metric
4EDL1S Linear 4A 1 Flange North American
4EDL1SC Linear 4A 1 Snap-In North American
4EDL1SM Linear 4A 1 Flange Metric
4EDL1SCM Linear 4A 1 Snap-In Metric
4EDL4 Linear 4A 4   Flange North American
4EDL4C Linear 4A 4   Snap-In North American
4EDL4M Linear 4A 4   Flange Metric
4EDL4CM Linear 4A 4   Snap-In Metric
6EDL1S Linear 6A 1 Flange North American
6EDL1SC Linear 6A 1 Snap-In North American
6EDL1SM Linear 6A 1 Flange Metric
6EDL1SCM Linear 6A 1 Snap-In Metric
6EDL4 Linear 6A 4   Flange North American
6EDL4C Linear 6A 4   Snap-In North American
6EDL4M Linear 6A 4   Flange Metric
6EDL4CM Linear 6A 4   Snap-In Metric
6EHL1S Medical 6A 1 Flange North American
6EHL1SC Medical 6A 1 Snap-In North American
6EHL1SM Medical 6A 1 Flange Metric
6EHL1SCM Medical 6A 1 Snap-In Metric
6EHL4 Medical 6A 4   Flange North American
6EHL4C Medical 6A 4   Snap-In North American
6EHL4M Medical 6A 4   Flange Metric
6EHL4CM Medical 6A 4   Snap-In Metric

• Includes DPST Switch
USA Customer Service: 1-800-468-2023
Corcom Products Direct: 1-847-680-7400
Germany: 49-89-6089-0
Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change.