Skip to content
 |  Tyco Electronics Segment Indicator  |   | 
Tyco Electronics - Our commitment. Your advantage.
Visit www.tycoelectronics.com Corcom Products
Home > Series > Medical > H

H Series

RFI Power Line Filters for UL2601 Healthcare Equipment

UL Recognized
CSA Certified
VDE Approved*

15EH4 - VDE rated at 10 amps, 250VAC

Catalog Data Sheet

H Series

The H series filters were developed specifically for health care applications. These products are suitable for uses in both patient care equipment and non-patient care equipment. Their extremely low leakage current allows compliance with UL2601 and CSA 22.2 health care specifications.

See Appendix C for more information on UL544 or UL2601 health care applications.

The H series features products with a variety of current ratings and case styles. Model 10EH4C additionally incorporates a separate ground circuit inductor to isolate equipment chassis from power line ground at RF frequencies.

Note: When using the models with an IEC connector, remember that the leakage current of the companion line cord, GA400, is more significant than that of the filter - which may disqualify this line cord from use in patient care applications.

Electrical Schematic

 

Resistor location for reference only

Specifications

Maximum leakage current, each line to ground:
  @ 120 VAC 60 Hz 2 µA
  @ 250 VAC 50 Hz 5 µA
     
Hipot rating (one minute):
  line to ground 2250 VDC
  line to line 1450 VDC
     
Operating Frequency: 50 / 60 Hz
     
Rated voltage (max.): 250 VAC
     
Rated Current: 3A to 15A *
  * refer to part number for rating
Minimum insertion loss in dB:
Line to ground in 50 ohm circuit

Current Rating Frequency - MHz
.15 .5 1 5 10 30

3A 18 27 30 30 27 18
6A 9 16 20 26 23 18
10A 7 13 15 17 16 14
15A 5 9 11 12 11 9


Case Styles

EH1

Typical Dimensions
Terminals (3) .250 Holes (2) .07 Dia.
6.35 1.8
Slot .07 x .16 Mounting Holes (2)
.188 Dia
1.8 x 4.1 4.78

EH3

 

Typical Dimensions
Wire Leads (min.) : 4.0 Wire Gauge: 18AWG
101.6
Mounting Holes (2)
.188 Dia    
4.78

 

EH4 & EH4C

Typical Dimensions
Terminals (3) .250 Holes (2) .07 Dia.
6.35 1.8
Slot .07 x .16 Mounting Holes (2)
.132 Dia
1.8 x 4.1 3.35
    with .260 x 90° countersunk for#4 Flathead Screw

EH5

EH8

Typical Dimensions
Wire Leads (min.) : 4.0 Wire Gauge: 18AWG
101.6
Mounting Holes (2)
.132 Dia    
3.35
with .260 x 90° countersunk for#4 Flathead Screw    

EH9

Typical Dimensions
Terminals (3) .250 Holes (2) .07 Dia.
6.35 1.8
Slot .07 x .16 Mounting Holes (2)
.132 Dia
1.8 x 4.1 3.35
    with .260 x 90° countersunk for#4 Flathead Screw

 

Recommended Panel Cutouts

tolerance ± .005 [0.13]

Note 1: EH4, EH4C and EH8 allow front or back mounting
Note 2: EH5 and EH9 allow back mounting only

Case Dimensions

Part No. A B C D E F
(max.) (max.) (max.) ±.015 (max.) (ref..)
± .38
3EH1, 6EH1,
10EH1
2.25 1.82 0.66 2.125 2.53  
57.2 46.1 16.7 53.98 64.2  

3EH3, 6EH3,
10EH3
.96 1.82 0.66 2.125 2.53  
24.40 46.1 16.7 53.98 64.2  

6EH4 2.20 1.19 0.81 1.575 1.98  
55.9 30.2 20.6 40.01 50.3  

6EH5 1.55 1.19 0.85 1.575 1.98 .295
39.4 30.2 21.6 40.01 50.3 7.5

6EH8 1.56 1.19 0.81 1.575 1.98 .295
39.7 30.2 20.6 40.01 50.3 7.5

6EH9 1.55 1.19 0.85 1.575 1.98  
39.4 30.2 21.6 40.01 50.3  

10EH4 2.62 1.19 0.81 1.575 1.98  
66.5 30.2 20.6 40.01 50.3  

10EH4C 2.62 1.19 0.81 1.575 1.98  
66.5 30.2 20.6 40.01 50.3  

15EH4 2.62 1.19 0.81 1.575 1.98  
66.5 30.2 20.6 40.01 50.3  

Part Numbers

Click on any part number for distributor stock status and eCommerce Links. For more information refer to the Corcom Part Numbering System reference.

3EH1 6EH8
3EH3 6EH9
6EH1 10EH1
6EH3 10EH3
6EH4 10EH4
6EH5 10EH4C
  15EH4
   
Line Cord GA400
Insulating Boot FA601

 

USA Customer Service: 1-800-468-2023
Corcom Products Direct: 1-847-680-7400
Germany: 49-89-6089-0
Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change.